DOI: http://dx.doi.org/10.18203/2320-6012.ijrms20192877

Evaluation of scatter radiation to the thyroid gland attributable to brain computed tomography scan in Port Harcourt, Nigeria

Ebbi D. Robinson, Christian C. Nzotta, Uwa Onwuchekwa

Abstract


Background: Computed tomography (CT) is an axial imaging modality that uses X-ray. The study is to determine the amount of X-radiation that reaches the thyroid glands during brain CT.

Methods: A prospective study was carried out in Rivers State University Teaching Hospital’s Radiology Department with 60 participants sent for brain CT scan. A 64 slice helical GE Optima CT machine was used while radiation dose reaching the thyroid gland was measured with themoluminiscent dosimeter (TLD) chips (TLD LiF-100). The TLD chip was placed on the anterior aspects of the neck at the level of 6th cervical vertebra (C6) and held in place with adhesive tapes. The TLD chip was later sent to the radiation dosimetric laboratory for reading. Collected data was analyzed using SPSS windows version 22.0 statistical software. A descriptive statistical tool was used to determine central tendencies. Pearson correlation and linear regression analysis models were also used to evaluate correlation between variables.

Results: The mean (±SD) CTDI, DLP and brain Effective Dose were 37.265±13.098mGy, 662.451±230.782mGy-cm and 1.667±0.603mSv respectively. The Scatter X-Radiation reaching the thyroid gland and resultant Effective Dose were 5.26±3.13mSv and 0.26±0.16mSv respectively. A positive correlation between brain and thyroid gland effective doses yielded a Pearson’s correlation coefficient (r) of 0.892 within a confidence interval of 0.01 (p value of 0.01) showing significant correlation.

Conclusions: The amount of radiation received by the thyroid gland during brain computed tomography scan is significant. Therefore, it is pertinent to protect the thyroid gland during the procedure.


Keywords


Brain Computed Tomography, Effective dose, Scatter radiation, Thyroid gland

Full Text:

PDF

References


Sinnott B, Ron E, Schneider AB. Exposing the Thyroid to Radiation: A Review of Its Current Extent, Risks, and Implications. Endocr Rev. 2010;31(5):756-73.

Aweda MA, Arogundade RA. Patient dose reduction methods in computerized tomography procedures: A review. Int J Phys Sci. 2007;2(1):1-9.

James E. Turner, Atoms, Radiation and Radiation Protection, 2nd Edition, John Wiley & Sons, Inc. 1995.

Ionising Radiation (Medical Exposure) Regulations (Statutory Instrument No.1059).; HMSO, London; 2000.

Bushberg JT, Seibert JA, Leidholdt EM, Boone JM, X-Ray Production, X-Ray Tubes, and Generators. In: Essentials of physics of medical imaging 2nd Ed. Philadelphia,USA: Lippincott willians and wilkins. 2002:97.

Hariqbal Singh. Physical Principle of CT Scan Imaging. In: Cross sectional imaging made easy. Jaapee Brothers Medical Publishers Ltd. 2011:3.

Mettler Jr FA, Wiest PW, Locken JA, Kelsey CA. CT scanning: patterns of use and dose. J Radiol Prot. 2000;20:353-9

Power SP, Moloney F, Twomey M, James K, O’Connor OJ, Maher MM. Computed tomography and patient risk: Facts, perceptions and uncertainties. Radiology. 2016;8(12):902-15.

National Research Council, Committee on the Biological Effects of Ionizing Radiation, Health Effects of Exposure to Low Levels of Ionizing Radiation, BEIR V, National Academy Press, Washington, D.C., 1990.

The-Evolution-Of-Ct-Scanner-Detectors. [Online] Available at: Http://Www.Medwow.Com/Articles/Ct-Scanner-2/The-Evolution-Of-Ct-Scanner-Detectors/.

Carina Storrs. How Much Do CT Scans Increase the Risk of Cancer? Scientific American July 1, 2013. Available at: https://www.scientificamerican.com/article/how-much-ct-scans-increase-risk-cancer/

Ryan S, McNicholas M, Eustace S. Head and Neck. In: Anatomy for Diagnostic Imaging. 3rd Ed. Philadelphia: Saunders Elsevier. 2011;233-234

Mazonakis M, Tzedakis A, Damilakis J, Gourtsoyiannis N. Thyroid dose from common head and neck CT examinations in children: is there an excess risk for thyroid cancer induction? Eur Radiol. 2007;17(5):1352-7.

International Atomic Energy Agency. Radiation protection of patients. Computed tomography. 2013.

Hart D, Wall BF. UK population dose from medical x-ray Examinations. Euro J Radiol. 2004;50(3):285-91.

Christner JA, Kofler JM, McCollough CH. Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection Publication 103 or dual-energy scanning. Am J Roentgenol. 2010;194(4):881-9.

Yamane, Taro. Statistics: An Introductory Analysis, 2nd Ed., New York: Harper and Row; 1967.

Abdullahi M, Shittu H, Joseph D, Aribisala A, Eshiett EP, Richard I, et al. Diagnostic reference level for adult brain computed tomography scans: A case study of a tertiary health care center in Nigeria. IOSR J Dental Med Sci. 2015;14(1):66-75.

Nzotta CC, Egbe N, Adejoh T, Nkubli BF, Ezeador IS. Radiation Dose from Exposure to Computed Tomography Scan of the Brain in A Reference Hospital In Nigeria. Pak J Radiol. 2016;26(2):77-80.

Ogbole GI, Obed R. Radiation doses in computed tomography: Need for optimization and application of dose reference levels in Nigeria. West Afr J Radiol. 2014;21(1):1-6

Adejoh T, Nzotta CC. Head computed tomography: Dose output and relationship with anthropotechnical parameters. West Afr J Radiol. 2016;23(2):113-7.

Chiegwu HU, Enweani IB, Nzotta CC, Ogbu SOI, Onyema AE, Ugwuanyi DC. Increasing radiation dose from Computed tomography versus Diagnostic Reference Levels: How prepared are we? Br J Med Medical Res. 2015;9(9):1-15.